Product Details:
Payment & Shipping Terms:
|
Model Numbe: | XB-OTS-UVA/B | Product Name: | UVA UVB Light Fastness Aging Test Chamber , UV Accelerated Weathering Tester |
---|---|---|---|
UV Wavelength: | 320-400um(UV-a), 280-320um(UV-B) | Lamp Power: | 40W/Unit, 8PCS For Unit Chamber |
Size Of Standard Specimen: | 75×150mm (customized Available) | Distance Between Lamps: | 70mm |
Usage: | UV Aging Test Chamber | Material: | Stainless Steel |
Internal Size: | 130*146*50cm | FOB Port: | Shenzhen |
Warranty: | 15 Months | ||
High Light: | uv accelerated weathering tester,uv aging test chamber |
UVA UVB Light Fastness Aging Test Chamber , UV Accelerated Weathering Tester
Product description:
Accelerated Weather UV test machine is applicable for the sunlight resistant test of nonmetallic materials and aging test of artificial light sources. A variety of industrial products can make the reliability test, and this product can simulate the product in the sun, rain, humidity and dew conditions, including damage caused by bleaching, color, brightness down, powder, crack, blur, brittle, intensity decreased and oxidation.
Chamber Structure:
The inner bladder of the chamber is made of imported high-class stainless steel plate while the outer bladder is made of SUS 304 steel plate.
The heating mode adopts inner bladder trough type heating with its rapid temperature rise and even temperature distribution.
Technical parameters: Various Material UV Lamp Weatherometer
Studio Size | 400W×1140D×390Hmm |
Outside dimension | 500W×1300D×1360Hmm |
Materials | Inside and Outside SUS#304 stainless steel |
Temperature range | RT+10~70°C |
Temperature uniformity | ±2°C |
Temperature fluctuation | ±0.5°C |
Temperature control | PID SSR control |
Humidity range: | ≥90%RH |
Controller | Korean TEMI 880 programmable controller |
Control mode | Balance temperature humidity control (BTHC) |
Test cycle setting | Illumination, condensation and water spray test cycle is programmable |
Lamp power | 40W/Piece |
Distance from sample to lamp | 50±2mm (adjustable) |
Centre distance between the lamp | 70mm |
Irradiance | 1.0W/cm2 |
Wavelength ultraviolet light | UV-A: 315-400nm; UV-B: 280-315nm (8pcs, 1600h lifetime) |
Effective irradiation areas | 640×1140mm |
Irradiation black panel temperature | 50°C~70°C |
Standard Specimen Size | 75×150mm or 75×3000mm (special specitications to be described in the contact) |
Required water depth for water channel | 25mm, automatically control |
Testing time | 0~999H, adjustable |
Power | 220V/50Hz /±10% 5KW |
Protection system | Overload short circuit protection, over temperature protection, water lackin |
Control System:
Adopts black aluminum plate to connect temperature sensor and adopts black board temperature meter to control heating so as to
ensure more stable temperature.
The radiometer probe is fixed so as to avoid the frequent installation and disassembly.
The illumination and condensation can be controlled independently or alternating and circularly.
Design criteria:
(1) ASTM D4587: Standard Practice for Fluorescent UVCondensation Exposures of Paint and Related Coatings
(2) ASTM D4799: Standard Practice for Accelerated Weathering Test Conditions and Procedures for Bituminous Materials
(Fluorescent UV, Water Spray, and Condensation Method)
(3) ISO 4892-1: Plastics – Methods of Exposure to Laboratory Light Sources – Part 1: General guidance
(4) ISO 4892-3: Plastics - Methods of Exposure to Laboratory Light Sources - Part 3: Fluorescent UV Lamps
(5) ASTM D4329: Standard Practice for Fluorescent UV Exposure of Plastics
(6) AATCC TM186: Weather Resistance: UV Light and Moisture Exposure
(7) ASTM D5208: Standard Practice for Fluorescent Ultraviolet (UV) Exposure of Photodegradable Plastics
(8) IEC 61345: UV Test for Photovaltalic (PV) Modules
Feature :
* Our UV aging test chamber series includes 4 different types of UV aging test boxes, QUV/se, QUV/basci, QUV/spray and QUV/cw. It can meet the testing standards of coatings, ink, adhesives, plastic products and other industries.
Standard:
1. ISO 48-3-2006, Plastics - Methods of exposure to laboratory light sources - Part 3: Fluorescent UV lamps;
2. ASTM D4329-2005, Standard Practice for Fluorescent UV Exposure of Plastics;
3. ASTM D4587-2005, Standard Practice for Fluorescent UV-Condensation Exposures of Paint and Related Coatings;
4. ASTM D5208-2001, Standard Practice for Fluorescent Ultraviolet (UV) Exposure of Photodegradable Plastics, etc.
Please feel FREE to contact us if you have any needs.
UVA UVB Light Fastness Aging Test Chamber , UV Accelerated Weathering Tester
Contact Person: Bonnie